Scaling of EDM calibration baselines by GPS and controlling of EDM parameters

Küçük Resim Yok

Tarih

2008

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

COMMONWEALTH ASSOC SURVEYING LAND ECONOMY, C A S L E

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

The calibration values of electromagnetic distance measurements, which have been given by their own firms, can lose their currency in time. So, the EDMs must be controlled in the particular time intervals. The EDM controls have been usually made in the EDM calibration baselines, which are constituted for this aim. Zero addition, scale coefficient and phase difference measurement function constitute the measurement function as a result of control done. The EDM calibration baseline must be scaled to determine the scale coefficient. Not only high accuracy instruments as Kern Mekometer 3000 but also GPS can be used for scaling. In this study, it has been investigated that possibility of scaling of EDM calibration baseline by using GPS and the scaled EDM calibration baseline, and the calibration parameters of Sokkisha SET2, Topcon GTS701, Topcon GTS 229 and Sokkia Power SET 2000 electronic tacheometers have also been estimated.

Açıklama

Anahtar Kelimeler

EDM calibration, baselines by GPS, EDM parameters

Kaynak

SURVEY REVIEW

WoS Q Değeri

Q3

Scopus Q Değeri

Q2

Cilt

40

Sayı

309