Right atrial perforation after an endocardial screw-in atrial lead implantation [Aktif atriyum elektrodu yerleştirilmesi sonrasi sa? atriyum delinmesi]

Küçük Resim Yok

Tarih

2013

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A 24-year-old female underwent implantation of a right-sided dual chamber permanent pacemaker for cardioinhibitory syncope with active fixation atrial lead. Five days after the procedure, the patient developed pleuritic chest pain and difficulty in breathing. Minimal pericardial effusion and right hemopneumothorax were found. The atrial active screw-in lead was visualized just above the right atrial appendage with its helix perforating the right atrial wall, pericardium and pleura, reaching the right mid-lobe. Lead extraction was performed, and a passive bipolar atrial lead was implanted during the same session without any problems. © 2013 Türk Kardiyoloji Derne?i.