dc.description.abstractSix unembalmed adult sheep mandibles were stripped of all soft tissues and sectioned at the midline. Each side had a sagittal split ramus osteotomy (SSRO) and was advanced 5 mm. Six of the hemimandibles were fixed with four-hole extended titanium miniplates and titanium screws, and the other six were fixed with four-hole extended absorbable plates and absorbable screws. At[ specimens were mounted in a servohydraulic testing unit, and a range of forces (0-140 N) was applied. Displacement of each proximal segment was recorded at 1 ON increments from 0 to 140 N. Values for the two groups were compared using the Mann-Whitney U-test, and significant differences in displacement were seen only at loads between 10 and 50 N. The results indicate that when absorbable miniptates are used intermaxillary fixation may be necessary to stabilise the bony fragments in the early postoperative period. (C) 2003 The British Association of Oral. and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.en_US
dc.identifier.doi10.1016/S0266-4356(03)00234-1en_US
dc.identifier.endpage132en_US
dc.identifier.issn0266-4356en_US
dc.identifier.issue2en_US
dc.identifier.pmid15013544en_US
dc.identifier.scopusqualityQ2en_US
dc.identifier.startpage127en_US
dc.identifier.urihttps://dx.doi.org/10.1016/S0266-4356(03)00234-1
dc.identifier.urihttps://hdl.handle.net/20.500.12395/18977
dc.identifier.volume42en_US
dc.identifier.wosWOS:000220502500006en_US
dc.identifier.wosqualityQ3en_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.indekslendigikaynakPubMeden_US
dc.language.isoenen_US
dc.publisherCHURCHILL LIVINGSTONEen_US
dc.relation.ispartofBRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERYen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.selcuk20240510_oaigen_US
dc.subjectsagittal split ramus osteotomyen_US
dc.subjecttitanium miniplateen_US
dc.subjecttitanium screwen_US
dc.subjectscrew fixationen_US
dc.titleComparison of stability of absorbable and titanium plate and screw fixation for sagittal split ramus osteotomyen_US
dc.typeArticleen_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
dolanmaz2004.pdf
Boyut:
192.13 KB
Biçim:
Adobe Portable Document Format
Açıklama: