Evaluation of Some Redox Mediators in the Design of Reagentless Amperometric Glucose Biosensor
Küçük Resim Yok
Tarih
2014
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
WILEY-V C H VERLAG GMBH
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
The influence of redox mediators (tetrathiafulvalene (TTF), tetrathiafulvalene-tetracyanoquinodimethane complex (TTF-TCNQ), phenazine methosulfate (PMS), 5,6-diamino-1,10-phenonthroline (5,6-DAP), potassium ferrocyanide (K4[Fe(CN)6]), methylene blue (MB) and toluidine blue (TB)) on the response of amperometric glucose biosensor was investigated in this research. Graphite rod electrode (GRE) modified by adsorbed mediator and glucose oxidase (GOx) cross-linked with glutaraldehyde was served as working electrode. The highest amperometric signals were observed by using TTF and TTF-TCNQ as mediators. The impact of amount of immobilized GOx, an optimal pH region for operation, stability and reproducibility of the analytic signal of designed biosensor using these mediators were investigated and discussed.
Açıklama
Anahtar Kelimeler
Amperometric glucose biosensor, Immobilized glucose oxidase, Electron transfer mediator, Methylene blue, Toluidine blue
Kaynak
ELECTROANALYSIS
WoS Q Değeri
Q2
Scopus Q Değeri
Q2
Cilt
26
Sayı
7