Comparison of Central Corneal Thickness Measurements With Optical Low-Coherence Reflectometry and Ultrasound Pachymetry and Reproducibility of Both Devices

Küçük Resim Yok

Tarih

2012

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

LIPPINCOTT WILLIAMS & WILKINS

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Purpose: To compare the accuracy of optical low-coherence reflectometry (Lenstar LS 900; Haag Streit, Koniz, Switzerland) and conventional ultrasound pachymetry (UP) (US 4000; Nidek, Japan) for measuring central corneal thickness and to assess the interexaminer reproducibilities of both devices. Methods: Department of Ophthalmology, Selcuklu Faculty of Medicine, Selcuk University was the study setting. Central corneal thickness (CCT) was measured in both eyes of 65 emmetropic patients by the same examiner using Lenstar LS 900 noncontact biometry and UP. To assess the interexaminer reproducibility, 2 different examiners consecutively obtained CCT measurements in 16 eyes with Lenstar LS 900 noncontact biometry and UP. Results: Mean CCT measurements with Lenstar LS 900 noncontact biometry and ultrasound biometry were 541.09 +/- 33.8 and 545.69 +/- 34.0, respectively. Significant correlation was present between these 2 measurements; intraclass correlation coefficient was found to be 95.8% (95% confidence interval: 93.2%-97.4%) (P < 0.001). Interexaminer reproducibility with Lenstar was high, and the intraclass correlation was found to be 99.3% (95% confidence interval: 98.0%-99.7%). We have also investigated the interexaminer reproducibility with UP in these 16 patients, and the results showed a 99.2% (95% confidence interval: 97.7%-99.7%) intraclass correlation. Conclusions: The results of Lenstar LS 900 noncontact biometry correlated with the results of UP. It may be used as an alternative method to UP for measuring CCT. Because of the mild underestimation of CCT by Lenstar, the measurements should be carefully evaluated in patients whose intraocular pressure must strictly be monitored.

Açıklama

29th Congress of the European-Society-of-Cataract-and-Refractive-Surgeons (ESCRS) -- SEP, 2011 -- Vienna, AUSTRIA

Anahtar Kelimeler

central corneal thickness measurement, optical low-coherence reflectometry, ultrasound pachymetry

Kaynak

CORNEA

WoS Q Değeri

Q2

Scopus Q Değeri

Q1

Cilt

31

Sayı

11

Künye