Biomechanical evaluation of malleable noncompression miniplates in mandibular angle fractures: an experimental study

dc.contributor.authorEsen, Alparslan
dc.contributor.authorDolanmaz, Dogan
dc.contributor.authorTuz, Hakan Hifzi
dc.date.accessioned2020-03-26T18:24:08Z
dc.date.available2020-03-26T18:24:08Z
dc.date.issued2012
dc.departmentSelçuk Üniversitesien_US
dc.description.abstractThe purpose of this experimental study was to test the reliability of a single malleable titanium miniplate using Champy's method of fixing fractures of the mandibular angle. Eighteen sheep hemimandibles were used to evaluate 2 plating techniques. The groups were tested with either a single non-compression titanium miniplate or a single malleable titanium miniplate. A cantilever bending biomechanical test model was used for the samples. Each group was tested with vertical forces using a servohydraulic testing unit. The displacement values in each group at each 10 N stage up to 90 N were compared using 2-way analysis of variance (ANOVA). The displacement values for the 2 groups differed significantly (p < 0.01). The variance analyses showed that the biomechanical behaviour of a single non-compression miniplate was better than that of a single malleable miniplate. The non-compression miniplate fixed by screws had greater resistance to occlusal loads than the malleable plate fixed by screws, and the malleable plate alone was not sufficient to withstand the early postoperative bite force. (C) 2011 The British Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.en_US
dc.identifier.doi10.1016/j.bjoms.2011.10.013en_US
dc.identifier.endpageE68en_US
dc.identifier.issn0266-4356en_US
dc.identifier.issn1532-1940en_US
dc.identifier.issue5en_US
dc.identifier.pmid22118919en_US
dc.identifier.scopusqualityQ2en_US
dc.identifier.startpageE65en_US
dc.identifier.urihttps://dx.doi.org/10.1016/j.bjoms.2011.10.013
dc.identifier.urihttps://hdl.handle.net/20.500.12395/27793
dc.identifier.volume50en_US
dc.identifier.wosWOS:000208858400001en_US
dc.identifier.wosqualityQ1en_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.indekslendigikaynakPubMeden_US
dc.language.isoenen_US
dc.publisherCHURCHILL LIVINGSTONEen_US
dc.relation.ispartofBRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERYen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.selcuk20240510_oaigen_US
dc.subjectMandibular angle fractureen_US
dc.subjectMalleable plateen_US
dc.subjectChampy methoden_US
dc.titleBiomechanical evaluation of malleable noncompression miniplates in mandibular angle fractures: an experimental studyen_US
dc.typeArticleen_US

Dosyalar