Evaluation of different miniplates in fixation of fractured human mandible with the finite element method

Küçük Resim Yok

Tarih

2007

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

MOSBY-ELSEVIER

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

The objective of this study was to develop a 3-dimensional finite element model ( FEM) to formulate biomechanical justification of the positioning of different plates to achieve stable fixation of a fractured mandible. Miniplate systems that give acceptable levels of rigidity were investigated, and recommendations about miniplate location, orientation, and type selection are made. A fracture near the body region was bridged with a variety of commonly used plate configurations. Number, positioning and type of the plate system parameters. The results of this fracture model support the advantage of 2-plate systems. Using a longer plate in the superior position and a shorter one in the inferior position produced a more stable condition. Number of screws or length of the miniplate had no significant effect on the stability of fractured segments. The results obtained from this study offer the choice of a particular plate size, thickness, design, or configuration for application and thus provide information for clinical use.